This topic contains information on the Fluke features supported and not supported by Smart Instrumentation.
Supported Features
-
Standard conventional 4-20ma transmitters such as flow, pressure, temperature, and electrical transducers.
-
Temperature transmitters: T/C and RTD (two wire RTD connections only).
-
Measuring frequency.
-
Loop power.
-
Square root extraction.
Refer to the following table for a complete list of physical source/measure capabilities:
mA |
VDC |
VAC |
W |
T/C |
RTD |
Pressure* |
HART |
Hz |
|
Sourcing |
Yes |
Yes |
No |
Yes |
Yes |
Yes |
Yes |
No |
No |
Measuring |
Yes |
Yes |
No |
Yes |
No |
Yes (2 wire only) |
Yes |
No |
Yes |
*Pressure calibration requires an additional module and pump.
Supported Pressure Units of Measure
Due to the limited number of base pressure units supported by Fluke, you can only use the following units of meassure when interfacing with Fluke:
-
Psi
-
Bar
-
InHg
-
mBar
-
inH2O
-
mH2O
-
ftH2O
-
Pa
-
kPa
-
gf/cm2 (Fluke: g/cm2)
Unsupported Fluke Features
In addition to the not supported features mentioned in the table in the previous section, the following Fluke features are not supported.
-
Custom units
-
HART communication
-
HART Scientific Drywell
-
Step and Ramp
-
Switch calibration
-
Delay option (temperature measurements).